SV 200A is the top-of-the range NOISE MONITORING STATION with built-in microphones for NOISE DIRECTIVITY detection. This revolutionary solution enables identification of dominant noise sources providing information about their location both in vertical and horizontal directions. In practice, the measurement of directionality gives the opportunity to indicate the dominant source of noise in the area of measurement or to exclude unwanted events.
Four additional microphones located on sides of the housing use the sound intensity technique to detect the DIRECTION of a DOMINANT NOISE SOURCE both in the vertical and horizontal axes. The Leq distribution in angle sectors is saved as the time-history and can be used for data filtering and reporting.
Station can perform a real-time frequency analysis in 1/1 and 1/3 OCTAVE bands and save it as time-history data. Additionally it can record the AUDIO SIGNAL for NOISE SOURCES RECOGNITION and data recalculation.
The ADVANCED ALARMS function can send e-mail and SMS notifications triggered by threshold level conditions combined with time conditions. Station’s status alarms are also available.
The 3G MODEM, WLAN and LAN provide fast data transfer over the Internet to PC with standard Internet connectivity. SvanNET enables a plug & play connection to Internet and easy management of measurement projects. Regardless of the SIM card type, Public or Private, SvanNET will establish connection, giving full access to the measurement data via WEB BROWSER. The Bluetooh® and Wireless LAN provide access point for an easy configuration with the SvanNET Application.
Following ISO 1996-2 requirements, the SV 200A is using the ELECTROSTATIC ACTUATOR to perform the periodic system check. CHECKING OF THE COMPLETE MEASUREMENT CHAIN including the microphone is the advantage of using electrostatic actuator method.
The accurate GPS module provides information on the localization as well as measurement TIME SYNCHRONIZATION.
Photos
Video
SV200A – Noise Directivity
Features
ACCURACY
- Class 1 IEC 61672-1: 2013 (type approved in EU)
NOISE SOURCES RECOGNITION:
- Four additional microphones located on sides of the housing to detect the DIRECTION of a DOMINANT NOISE SOURCE both in the vertical and horizontal axes
- Real-time frequency analysis in 1/1 and 1/3 OCTAVE bands and
- AUDIO SIGNAL for listening and recalculation
REMOTE COMMUNICATION
- The ADVANCED ALARMS function can send e-mail and SMS notifications
- 3G/4G MODEM, WLAN and LAN , Bluetooth
SYSTEM CHECK
- ELECTROSTATIC ACTUATOR to perform the periodic system check
- Status alarms on battery, powering, memory, measurement status
- The accurate GPS module provides information on the localization as well as
measurement TIME SYNCHRONIZATION
Specification
SV 200A Sound Level Meter Specification
Standards | Class 1: IEC 61672-1:2013, Class 1: IEC 61260:2014 |
Number of Profiles | 3 |
Weighting Filters | A, C, Z |
Time constants | Slow, Fast, Impulse |
RMS Detector | Digital True RMS detector with Peak detection, resolution 0.1 dB |
Microphone | Microtech Gefell MK 255, 50 mV/Pa, prepolarised 1/2” condenser microphone |
Preamplifier | Integrated |
Linear Operating Range | 25 dBA RMS ÷ 133 dBA Peak (in accordance to IEC 61672) |
Total Dynamic Measurement Range | 15 dBA RMS ÷ 133 dBA Peak (typical from noise floor to the maximum level) |
Internal Noise Level | less than 15 dBA RMS |
Dynamic Range | >115 dB |
Frequency Range | 3.5 Hz ÷ 20 kHz |
Meter Mode Results | Elapsed time, Lxy (SPL), Lxeq (LEQ), Lxpeak (PEAK), Lxymax (MAX), Lxymin (MIN), Lxye (SEL), LN (LEQ STATISTICS), Lden, LEPd, Ltm3, Ltm5 Simultaneous measurement in three profiles with independent set of filters (x) and detectors (y) |
Statistics | Ln (L1-L99), complete histogram in meter mode
Ln (L1-L99) statistics in 1/1 or 1/3 octaves |
Data Logger | logging of summary results with step from 1 s
logging of time history with step from 20 ms |
Audio Recording (option) | Audio events recording during measurement, trigger or continuous mode, 12 / 24 / 48 kHz sampling rate, 16 or 24 bit, wav format |
Sound Analyser |
|
1/1 Octave Analysis | Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 31.5 Hz to 16 kHz |
1/3 Octave Analysis | Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 20 Hz to 20 kHz |
General information |
|
Ingress Protection Rating | IP 65 |
Inputs | Power supply LEMO 3-pin, extended I/O port LEMO 9-pin |
Remote Calibration | Built-in electrostatic actuator, triggered manually or in automated mode |
Memory | 16 GB |
Display & Keyboard | 1.1” OLED colour display, 5 push-buttons |
Communication Interfaces | USB, RS 232, UART (TTL), LAN, Bluetooth®, 3G modem, WLAN |
Power supply | Li-Ion rechargeable battery (non-removable) |
Solar Panel (not included) | |
SB274 AC power supply (included) Input 100 ÷ 240 VAC, output +15 VDC 2.67 A, IP 67 housing | |
External DC source (not included) – voltage range 10.5 V – 24 V | |
Environmental Conditions | Temperature from -30 oC* to 70 oC
*Only with external powering otherwise -10oC to 50 oC |
Humidity up to 99% RH, non-condensed | |
Dimensions | 840 mm length; 70 mm diameter excluding windscreen; windscreen diameter 130 mm |
Weight | Approx.3.2 kg |
*Continuous product development and innovation are the policy of our company. Therefore, we reserve the right to change the specifications without prior notice.